Low Frequency Conducted Immunity Test System

No. produk

Tinggalkan pesan

=

Low Frequency Conducted Immunity Test System

Tinggalkan pesan

=
Deskripsi Produk

LFCI61000-1619 Low Frequency Conducted Immunity Test System with various testing accessories such as co-anti-disturbance test devices, coupling removal network, and current injection device can meet different test needs. The low-frequency common mode/differential mode conducted immunity testing system is capable of simulating the fundamental frequency, major harmonics, and interharmonics of power distribution systems or power electronic devices (such as power converters). The interference can be injected into the grounding system through stray capacitance or coupled onto the signal and control lines through induction.

Standar:
IEC/EN 61000-4-19 (GB/T17626.19) “Ectromagnetic compatibility (EMC) – Part 4-19: Testing and measurement techniques – Test for immunity to conducted, differential mode disturbances and signalling in the frequency range 2 kHz to 150 kHz at a.c. power ports”
IEC/EN 61000-4-16 (GB/T17626.16) “Ectromagnetic compatibility (EMC) – Part 4-16: Testing and measurement techniques – Test for immunity to conducted, common mode disturbances in the frequency range 0 Hz to 150 kHz”
JJF 1245.4-2019 “Outline of Type Evaluation of Installed AC Energy Meters-Special Requirements and Safety Requirements”
IEC 62052-11 (GB/T17215.211) “electricity metering equipment – General requirements, tests and test conditions – Part 11: Metering equipment”
IEC/EN 60255-22-7 (GB/T 14598.19) “Electrical relays – Part 22-7: Electrical disturbance tests for measuring relays and protection equipment – Power frequency immunity tests”

Aplication:
Intelligent meters utilize PLC data transmission for smart metering; touch dimming lights; communication systems (such as Ethernet systems, ISDN, ADSL modems, IP network branch switches, routers); contactless card readers, credit card terminals; alarm systems; electronic controls (such as heating systems, streetlights, traffic lights); household appliances (such as steam irons, coffee machines, microwaves, washing machines), etc.

Konfigurasi:

Model LISUNProduct nameTechnical specification
LFCI61000-1619Low Frequency Conducted
Immunity Test System
Compliant with the requirements of IEC61000-4-16 and IEC61000-4-19 testing standards, it features built-in signal generator, power amplifier, power meter and attenuator. The testing frequency range is DC to 150kHz, with voltage output ranging from 0.1V to 30V and current output ranging from 0.1A to 6A. The output impedance is 50 Ω±10%, and it also includes pulse modulation capability.
LFCI61000-16STLow Frequency Common
Mode Conducted Immunity Test Device
It works together with host LFCI61000-1619 is compatible with IEC 61000-4-16 standard for testing voltage in the range of 30-330V, with short-term tests conducted at frequencies of DC, 16²⁄³ Hz, 50Hz, and 60Hz.
CN-M2-16-DCCoupling Network
(common mode)
Compliant with IEC 61000-4-16 standard; suitable for EUT power supply DC 230V/16A (single-phase L/N); maximum test level for continuous testing at 35V and for short-time testing at 350V.
CN-M4-32Coupling Network
(common mode)
Compliant with the IEC 61000-4-16 standard, suitable for the EUT power supply of AC 440V/32A (three-phase four-wire, with switchable lines 2, 3, and 4). The highest test level for continuous testing is 35V, and for short-duration testing is 350V.
LFCI61000-19INCurrent Injection Device
(differential mode)
Compliant with the standard IEC 61000-4-19 (current method); frequency range: 2kHz to 150kHz; EUT power supply: maximum injected current 6A; three-phase switchable.
CDND-M3-16Coupling/Decoupling
Network (differential mode)
Compliant with standard IEC 61000-4-19 (Voltage Method); frequency range 2kHz~150kHz. EUT power supply: AC 250V (single phase), AC 440V (three phase), and DC 500V, with a maximum current of 16A.
CAL-CDND19CDND-M3-16 calibration
part
Compliant with the standard IEC61000-4-19 (voltage method); equipped with two sets of Balun converters and adapters.
LFCI61000-V1.0Perangkat Lunak KontrolTo control LFCI61000-1619 and other optional accessories for testing purposes, this software is provided for free and can be installed on Win7, Win8, Win10, Win11.

Typic Appications 1: LFCI61000-1619, LFCI61000-19IN and LFCI61000-V1.0 are fully compliant with IEC 62052-11 (GB/T17215.211), which is a special standard for the electric energy meter industry.

Typic Appications 2: The LFCI61000-1619, LFCI61000-19IN, CDND-M3-16, CAL-CDND19 and LFCI61000-V1.0 completely comply with the differential mode voltage test and differential mode current test requirements of IEC/EN 61000-4-19 (GB/T17626.19).

106B9W44TC0Z883GYN

Table 1 – Test levels in the 2kHz to 150kHz frequency range for differential voltage testing

2

Table 2-Test levels in the 2kHz to 150kHz frequency range for differential current testing

IEC/EN 61000-4-19 Clause 5.1 Differential Model Voltage Test Technical Parameters

Rentang Frekuensi2 kHz~150 kHz
WaveformSine wave, with total harmonic distortion below 5%
Opening Voltage0.1 V~25 V±5%
Pulse Modulation3Hz~1kHz
Pulse Modulation Occupation Ratio50%
Output Impedance:10 Ω±30%

IEC/EN 61000-4-19 Clause 5.2 Differential Model Current Test Technical Parameters

Rentang Frekuensi2 kHz~150 kHz
Max. Current7 A
Load Output Impedance1Ω±0.3 Ω
Pulse Modulation3Hz~1kHz
Duty Ratio50%

Typic Appications 3: LFCI61000-1619, LFCI61000-16ST, CN-M2-16-DC (or CN-M4-32) and LFCI61000-V1.0 fully comply with the requirements of IEC/EN 61000-4-16 (GB/T17626.16) for continuous disturbance, short-term interruption disturbance, and testing within the frequency range of 15Hz to 150kHz.

3

Test levels for continuous disturbances at the electricity supply frequency(Table 1 of EN61000-4-16)

4

Test levels for short duration disturbances at the electricity supply frequency(Table 2 of EN61000-4-16)

5

Test levels for continuous disturbances in the frequency range 15Hz to 150kHz (Table 3 of EN61000-4-16)

IEC/EN 61000-4-16 Table 1 Levels for continuous disturbance Technical Parameters:

Test FrequencyDC, 16²⁄³ Hz,50 Hz and 60 Hz
Opening Voltage1 V(-10%)~50V(+10%)
Impedansi Keluaran50 Ω±10%

IEC/EN 61000-4-16 Table 2 Level for short duration disturbance Technical Parameters:

Test FrequencyDC, 16²⁄³ Hz,50 Hz and 60 Hz
Opening Voltage10(-10%)V~330(+10%)V
Impedansi Keluaran50 Ω±10%
Down/Rise TimeBetween 1μs~5μs(DC)
The toggle between ON/OFF of the output voltageSwitching at a phase angle of 0 degrees (0°± 5%) (16²⁄³ Hz, 50 Hz, and 60 Hz).

IEC/EN 61000-4-16 Table 3 Test Level in the frequency range 15Hz to 150kHz Technical Parameters:

Rentang Frekuensi15 Hz~300 kHz
Opening Volatge0.1V(-10%)~50V(+10%)
Impedansi Keluaran50 Ω±10%
Total Harmonic Distortion<1%(Sine Wave)