LMS-9000C High Precision CCD Spectroradiometer is according to LM-79 clause 9.1. It is designed for any visible optical spectrum distribution analysis, mimicking various light sources in the visible region by feedback control of the radiant power emitted by individual LEDs. This LED optical spectrum analyzer is used as a transfer standard for photometric, colorimetric and radiometric applications.

LMS-9000C can do quick test with high accuracy. It used in scientific lab. It works with 積分球 and standard lamp as a system: LPCE-2 High Precision Spectroradiometer Integrating Sphere System.

スタンダード:
CIE 177 "白色LED光源の演色性"
CIE84 "光束の測定"
CIE-13.3 "光源の演色特性の測定方法および指定方法"
LM-79-19 「固体照明製品の光学的および電気的測定」
IESNA LM-63-2 "測光データと関連情報の電子転送のためのIESNA標準ファイルフォーマット" (IESNA Standard File Format for Electronic Transfer of Photometric Data and Related Information)
IES LM-80-08 "LED光源のルーメン維持率の測定"
ANSI-C78.377 "電気ランプ用固体照明製品の色度に関する規格"
(EU) 2019/2015 「委員会委任規則

測定:
- 測色色度座標、CCT、色比、ピーク波長、半値幅、優勢波長、色純度、CRI、CQS、TM-30(Rf、Rg)、スペクトルテスト
- 測光光束、光効率、放射電力、WPE、EQE、EEI、エネルギー効率クラス、瞳束、瞳束効率、瞳係数、サイトピックフラックス、植物成長ランプPARおよびPPF
• Electrical: Voltage, Current, Power, Power Factor, Displacement Factor, Harmonic
• LED optical maintenance test: Flux VS time, CCT VS time, CRI VS time, Power VS time, Power Factor VS time, Current VS time and Flux Efficiency VS time

仕様:
- 分光波長精度:±0.3nm、波長再現性:±0.1nm
- サンプルのスキャンステップ±0.1nm
- 色度座標(Δx、Δy)の精度:±0.002(標準Aランプ使用時)
- 相関色温度CCT:1,500K〜100,000K、CCT精度:±0.3%
- 演色評価数範囲:0~100.0、精度:±(0.3%rd±0.3)
- 光束範囲0.01-200,000lm; 測光線形正確さ: ±0.5%
- 迷光<0.015%(600nm)および<0.03%(435nm)
- 積分時間: 0.1~10,000ms
- 積分球の内側と外側の温度を測定できる。
• Total flux testing method: spectrum, photometric and spectrum with photometric revision

LISUNモデル LMS-9000C LMS-9000CUVIS LMS-9000CVIS-NIR LMS-9000CUV
波長 350-800nm 200-800nm 350-1050nm 200-400nm

The LMS-9000C (which is updated on LMS-9000B) uses the band pass-filter wheel correcting technique, spectrometer & broadband-radiometer & photometer combined technique, and modified NIST stray light correction technology. The LMS-9000C spectroradiometer can realize ultra low stray light and super photometry linearity in overall dynamic range.