WKS-1010 Mask Bacterial Filtration Efficiency (BFE) Tester is used to test the percentage of the mask material to filter out the suspended particles containing bacteria under the specified flow rate. It is suitable for the performance test of the bacterial filtration efficiency of medical surgical masks by metrological inspection departments, scientific research institutes, medical mask manufacturers and other related departments.

Standard:
ASTM F2100 “Standard Specification for Performance of Materials Used in Medical Face Masks”
YY0469-2011, YY/T 0969-2013

Specification:

Main parameter Parameter range Resolution Accuracy
Sampling flow 28.3L/min 0.1L/min Better than ±2.5%
Spray flow (8~10)L/min 0.1L/min Better than ±2.5%
Peristaltic pump flow (0.006~3.0)ml/min 0.001ml/min Better than ±2.5%
Front pressure of flowmeter (-20~0)KPa 0.01KPa Better than ±2.5%
Front pressure of Spray flowmeter (0~300)KPa 0.1KPa Better than ±2.5%
Ambient temperature (-40~99)℃ 0.1℃ Better than ±2.5%
Negative pressure of aerosol chamber (-90~-120)Pa 0.1Pa Better than ±2.0%
Negative pressure of cabinet -50~-200Pa
Data storage capacity >100000 groups
Specification and quantity of vortex mixer test tube Φ16×150mm tube;8
High efficiency air filter characteristics Filtration efficiency for particles above 0.3um≥99.99%
Mass median diameter of aerosol generator 3.0± 0.3μm,Geometric standard deviation≤1.5
Integrated 6 Grades Andersen

Sampler capture particle size

Grade I>7μm,Grade Ⅱ(4.7~7)μm,Grade Ⅲ(3.3~4.7)μm,Grade Ⅳ(2.1~3.3)μm,Grade Ⅴ(1.1~2.1)μm,Grade Ⅵ(0.6~1.1)μm
Aerosol chamber specifications 60cm(Length)×8.5 cm(Diameter)× 3cm(Thickness)
Total number of particles of positive quality control sampler 2200±500cfu
Ventilation flow of negative pressure cabinet ≥5m3/min
Size of negative pressure cabinet 755×400×990mm (W×D×H)
Size of host 1200×630×2100mm (W×D×H)
Instrument noise <65dB(A)
Total Weight About150kg
Power supply AC220V±10%,50Hz
Power consumption <1500W

Features:

  • Negative pressure experiment system to ensure the safety of operators;
  • Negative pressure cabinet built-in peristaltic pump, integrated 6 Grades Andersen;
  • The flow rate of the peristaltic pump can be set;
  • The special microbial aerosol generator bacterial liquid spray flow rate can be set, and the atomization effect is good;
  • Embedded high-speed industrial microcomputer control;
  • 10.4 inch industrial grade high-brightness color touch screen;
  • USB interface, support U disk data transfer;
  • Built-in high-brightness lighting lamp in the cabinet;
  • Built-in leakage protection switch to protect the safety of operators;
  • Heat insulation and flame retardant between the inner and outer layers;
  • The front opening and closing glass door is convenient for the experimenter to observe and operate;